6332D Airflow Transducer Specifications
Measurement Object
Clean air & compatible, non combustible gas
MeasurementRange
Max. range .1 - 50 meters / sec. (Range dependent on probe type)
Temperature Compensation Range
5-40 deg. C
40-80 deg. C
.1 to 4.99 meters / sec.
±.25 meters / sec.
±.35 meters / sec.
5.00 to 9.99 meters / sec.
±.50 meters / sec.
±.70 meters / sec.
10.0 to 24.9 meters / sec.
±1.25 meters / sec.
±1.75 meters / sec.
±2.50 meters / sec.
±3.50 meters / sec.
Selectable Range Option
(Selectable w/ Rotary Switch)
0 to 2, 0 to 5, 0 to 10, 0 to 25, or 0 to 50 meters / sec.
Velocity Ranges - 0-2, 0-5, 0-10 meters / sec. ? Display Resolution: .01 meters / sec.Velocity Ranges - 0-25, 0-50 meters / sec. ? Display Resolution: .1 meters / sec.
DC 12 to 24 Volts
approximately 2.0 Watts (with these conditions - supplied power 12 Volts, air velocity - approximately 10 meters / sec., ustilizing unit w/ display and probe 0965-03)Rate of power consumption changes based on conditions like probe type, air velocity & use of a display
Connection Wire Size
.5 - 1.5 millimeter2
Operating Temperature Range
5 - 40 deg. C
Storage Temperature Range
-10 - 50 deg. C
Dimensions
approximately 128(W)×78(D)×30(H) millimeter
Weight
approximately 320 gram (11oz.)
Standard Accessory
Operation Manual : 1, 1 : Main Unit Case
Optional Accessory
Probe Cable (10 meter, 20 meter, 30 meter), Display Unit (Includes case w/ display window & 2 installation screws), AC Adapter DC 12 Volts)
Probe Specification
Model Number
Measurement Range
Sensor
0962-00, 0963-00
.1-50.0 meters / sec.
Uni directional
0964-01, 0964-02
(Needle) Omni directional
0965-00, 0965-01
.1-25.0 meters / sec.
(Spherical) Omni directional
0965-03, 0965-04
Mini temp. compensation sensor integrated type
0965-07, 0965-08
Mini-temperature-compensation-sensor independent type
Copyright Green Tech Tools. All Rights Reserved.