The HD 150 uses a capacitive sensing element to make relative humidity measurements. A capacitive polymer layer inside the probe reacts with humidity which occurs between 2 metal layers that cover a glass substract. Water absorption, which is a function of (RH) relative humidity in the immediate environment changes the dielectric constant. The signal measured is directly proportionnal with relative humidity & is dependent on atmospheric pressure.
Kimo HD 150 Technical Features
Measuring element
Hygrometry and temp. :CMOS sensor
Power supply
24 Vac/Vdc +/- 10 percent
Display
2 lines, LCD technology. Sizes 50 x 34.9 mm.1 line of 5 digits with 7 segments (value)1 line of 5 digits with 16 segments (unit)
Housing
Shock-proof made of ABS, IP54 protectionor IP67 with food industry protective cover
Key pad
5 keys, Metal coating
Cable
Retractable, length 450 mm, up - 2.4 m whenreleased (HD 100)Straight, length 1 meter (HD 150)
Conformity
electro-magnetic compatibility (NF-EN-61326-1 guidelines)
Power
1 alkaline 9 Volt battery 6LR61
Operating temp. (instrument)
from 0 - 50 deg. C
Operating temp. (probe)
from -20 - 70 deg. C
Storage temp.
from -20 - 80 deg. C
Adjustable auto shut-off
0 - 120 min
Weight.
190g
Languages.
French, english
Kimo HD 150 Specifications
Measuring Units
Measuring Ranges
Accuracy*
Resolutions
RELATIVE HUMIDITY (RH)
percentRH
from 5 - 95 percentRH
Accuracy* (Repeatability, linearity, hysteresis) : +/- 1.8 percentRH (from 15 deg. C - 25 deg. C)Factory calibration incertainty : +/- 0.88 percentRHTemperature dependence :
+/- 0.04 x (T-20) percentRH (if T<15 deg. C or T>25 deg. C)
0.1 percentRH
DEW POINT
deg. C td, deg. F td
from -40 - +70 deg. Ctd
+/- 0.8 percent of reading +/- 0.6 deg. Ctd
0.1 deg. Ctd
AMBIENT TEMPERATURE
deg. C, deg. F
from -20 - +70 deg. C
+/- 0.4 percent of reading +/- 0.3 deg. C
0.1 deg. C