AQ 200 Probe Specifications |
| Measuring Units | Measuring Ranges | Accuracy | Resolutions |
CURRENT / VOLTAGE |
| V, mA | from 0 to 2,5 V from 0 to 10 V from 0 to 4/20 mA | +/- 2mV +/- 10mV +/- 0.01mA | 0.001 V 0.01 V 0.01 mA |
THERMOCOUPLE (See related datasheet) |
| deg. C, deg. F | K : from -200 to 1,300 deg. C J : from -100 to 750 deg. C T : from -200 to 400 deg. C | +/- 1,1 deg. C or +/- 0,4 percent of reading +/- 0,8 deg. C or +/- 0,4 percent of reading +/- 0,5 deg. C or +/- 0,4 percent of reading | 0.1 deg. C 0.1 deg. C 0.1 deg. C |
Climatic conditions |
Hygro. | percentRH | From 5 to 95 percentHR | Accuracy** (Repeatability, linearity,hysteresis) :+/- 1,8 percentRH (from 15 deg. C to 25 deg. C) Factory calibration uncertainty: +/- 0,88 percentRH Temperature dependence : +/- 0.04 x (T-20) percentRH (if T<15 deg. C or T>25 deg. C) | 0.1 percentRH |
Temp.
Air pressure | deg. C, deg. F
hPa | From -20 to +80 deg. C
From 800 to 1100 hPa | +/- 0.4 percent of reading +/- 0.3 deg. C
+/- 3 hPa | 0,1 deg. C
1 hPa |
CO / Temperature |
Temp.
CO | deg. C, deg. F
ppm | From -20 to +80 deg. C
From 0 to 200 ppm From 0 to 5000 ppm | +/- 0.3 percent of reading +/- 0.25 deg. C
+/- 3 ppm
+/- 1.5 percent of reading | 0,1 deg. C
0,1 ppm |
CO2 / Temperature |
Temp.
CO2 | deg. C, deg. F
ppm | From -20 to +80 deg. C From 0 to 5000 ppm | +/- 0.3 percent of reading +/- 0.25 deg. C
+/- 3 percent of reading or +/- 50ppm | 0,1 deg. C
1 ppm |
CO2 / Temperature/ Hygrometry |
Temp.
CO2 Hygro. | deg. C, deg. F
ppm
percentRH | From -20 to +80 deg. C From 0 to 5000 ppm From 5 to 95 percentHR | +/- 0.3 percent of reading +/- 0.25 deg. C
+/- 3 percent of reading or +/- 50ppm
See Climatic conditions module | 0,1 deg. C
1 ppm
0,1 percentRH |
Omnidirectionnal probe (Air velocity / Temperature / Hygrometry) |
Air velocity
Temp. Hygro. | m/s
deg. C, deg. F
percentRH | From 0 to 5 m/s From -20 to +80 deg. C From 5 to 95 percentHR | +/- 3 percent of reading +/- 0.05 m/s
+/- 0.3 percent of reading +/- 0.25 deg. C
See Climatic conditions module | 0,01 m/s
0,1 deg. C
0,1 percentRH |
Pt100 Smart-plus or wireless probes (See related datasheet) |
| deg. C, deg. F | From -50 to 250 deg. C (According to model) | +/- 0,3 percent of reading +/- 0.25 deg. C (According to model) | 0,01 deg. C |