The HD 200 HRS uses a capacitive sensing element to make relative humidity measurements. A capacitive polymer layer inside the probe reacts with humidity which occurs between 2 metal layers that cover a glass substract. Water absorption, which is a function of (RH) relative humidity in the immediate environment changes the dielectric constant. The signal measured is directly proportional with relative humidity & is dependent on atmospheric pressure.
Kimo HD 200 HRS Technical Features
Sensing elements
Hygrometry : capacitive sensor
Temperature : Pt100 1/3 DIN
HD200 connection
On the top : 2 secured mini-DIN connectors for SMART-Plus probes
Left side : 1 USB port for KIMO cable only1 power supply plug
Interchangeable measurement modules
Thermocouple module :
Connection :4 inputs for compensated miniature maleplug of thermocouple type K, J or T Class 1(as per IEC 584-3 norm)
Current / voltage module :Connection : 2 stereo jacks
Display
Graphic display 128x128 pixels
Dim. 50 x 54 mm
Blue backlight
Display of 6 measurements (including 4 simultaneously)
Housing
ABS shock-proof
IP54
Keypad
5 keys - Metal coated
joystick
Conformity
Electro magnetic compatibility
(NF-EN-61326-1 norm)
Power
4 alkaline 1.5 Volt LR6 batteries
Operating environment
Neutral gas
Operating temp.
from 0 - 50 deg. C
Storage temp.
from -20 - 80 deg. C
Auto shut-off
adjustable from 0 - 120 min
Weight
340g
Languages
French, English, Dutch, German, Italian,Spanish, Portuguese, Swedish, Norwegian, Finn, Danish
Kimo HD 200 HRS Specifications
Measuring Units
Measuring Ranges
Accuracy*
Resolutions
CURRENT / VOLTAGE
V, mA
from 0 - 2,5 Vfrom 0 - 10 Vfrom 0 - 4/20 mA
±2mV±10mV±0.01mA
0.001 V0.01 V0.01 mA
THERMOCOUPLE (See related datasheet)
deg. C, deg. F
K : from -200 - 1,300 deg. CJ : from -100 - 750 deg. CT : from -200 - 400 deg. C
±1,1 deg. C or ±0,4% of reading±0,8 deg. C or ±0,4% of reading±0,5 deg. C or ±0,4% of reading
0.1 deg. C0.1 deg. C0.1 deg. C
HYGROMETRY PROBES
Relative humidityAbsolute humidity
%RHg/Kg
From 3 - 98 %RHFrom 0 - 600 g/Kg
Accuracy** (Repeatability, linearity,hysteresis) :±1,5%RH (from 15 deg. C - 25 deg. C)Factory calibration uncertainty: ±0,88 %RHTemperature dependence : ±0.04 x (T-20) %RH (if T<15 deg. C or T>25 deg. C)
0.1 %RH0.1 g/Kg
Dew pointAmbient temp.
deg. C td , deg. Ft td deg. C, deg. F
From -50 - +80 deg. Ctd From -20 - +80 deg. C
±0.6% of reading ±0.5 deg. C±0.3% of reading ±0.25 deg. C
From -50 - +80 deg. Ctd From -40 - +180 deg. C